Gas Cylinder

Ìý

Transport

A:  Locate your gas cylinder from the gas cylinder cage and transport
Step 1: Locate your gas cylinder from the gas cylinder cage and transport to the lab with the use of a gas cylinder cart.
B:  Make sure the cylinder is properly secured to the cart before
Step 2: Make sure the cylinder is properly secured to the cart before transporting.
C:  Secure the cylinder in the lab, to the wall or bench with a
Step 3: Secure the cylinder in the lab, to the wall or bench with a gas cylinder clamp. Important: 1 metal chain for 1 gas cylinder. Do not secure multiple gas cylinders with 1 chain. Have a minimum distance of 1.5 m between two incompatible gases such as H2 and Air.

Connect gas regulator

A:  Undo the cap and screw on the appropriate regulator for the gas.
Step 4: Undo the cap and screw on the appropriate regulator for the gas.
B 5: Use a proper regulator wrench to tighten the regulator.
Step 5: Use a proper regulator wrench to tighten the regulator. Do not over tighten and do not use adjustable wrench. Do not use Teflon tape on the cylinder connections.
C: Check that the flow control valve is closed.
Step 6: Check that the flow control valve is closed.

Check pressure and valves

Step 7:  Turn the pressure adjusting screw on the regulator
Step 7: Turn the pressure adjusting screw on the regulator counterclockwise until it turns freely.
A:  Slowly open the cylinder valve. A ½ turn is usually sufficient.
Step 8: Slowly open the cylinder valve. One full turn is usually sufficient.
​
C: Check the gas cylinder pressure gauge.  A full cylinder has
Step 9: Check the gas cylinder pressure gauge. A full cylinder has approx. 2200 psi except for acetylene, carbon dioxide or special gas mixtures.

Check pressure and leak test

Step 10:  Turn the pressure adjusting screw clockwise until the
Step 10: Turn the pressure adjusting screw clockwise until the required pressure is reached (gauge further from the gas cylinder valve).
Step 11:  Check for gas leaks by using Snoop(TM) on all the threaded
Step 11: Check for gas leaks by using Snoop(TM) on all the threaded connections including the cylinder valve.
Step 12: If all is correct, control the gas flow by the flow control valve
Step 12: If all is correct, control the gas flow by the flow control valve on the regulator. Do not use the pressure adjusting screw for flow control.

Shut down

Step 13: For shut down: shut the cylinder valve, allow residual gases
Step 13: For shut down: shut the cylinder valve, allow residual gases in valve and lines to drain.
Step 14: When gauges fall to zero, turn the pressure adjusting
Step 14: When gauges fall to zero, turn the pressure adjusting screw counter-clockwise until it runs freely and close the flow control valve.
Step 15: When moving the gas cylinder always remove the
Step 15: When moving the gas cylinder always remove the regulator, replace the cap, and then move cylinder using a gas cylinder cart.